@ultrasoundtok: #Ultrasound #VascularUltrasound #Sonography #DopplerPhysics #Sonographer

UltraTok
UltraTok
Open In TikTok:
Region: US
Monday 21 September 2026 00:37:27 GMT
31676
720
56
14

Music

Download

Comments

christys16
Christy Seabolt Crisp :
That angle is catastrophic!
2026-09-21 08:59:03
4
garcia.miranda
Miranda Garcia :
Rock the probe & turn the gain down
2026-09-21 04:19:59
16
hfontenot729
Heather Schoeffler F :
Steering and angle needs to be positioned parallel to vessels walls
2026-09-21 02:36:38
56
mcsmallkyle
Landen Distler :
I mean in all actuality its a fine image you can always make it picture perfect but if you understand waveform analysis then you'll be fine
2026-09-21 04:26:15
11
72kmbrly
Kimmie :
Gain is too high
2026-09-22 22:36:10
4
kingrt5
KingRayHenry :
The box is cutting off the vessel, the angle isn’t parallel, pulse wave Doppler needs to be turned down a little
2026-09-21 01:51:23
6
marcoc5
Marco C :
gain and box the wrong angle
2026-10-14 00:23:17
1
user6001981934879
Bennamic :
argumentation
2026-10-07 18:54:27
0
roadrunneress_318
Betsy :
Adjust the gain. Clean up artifact.
2026-10-13 09:05:47
1
___deemarie
Dee Marie :
Need to dive that vessel! Correct angle and steer
2026-09-21 05:26:21
2
theerickmode
Erick_Zxav :
Heel toe technique was necessary and then steer the box to the direction of flow before applying power which of course comes at an angle 📐 to the vessel smthg like less than 60 degrees, a 45 may be
2026-10-10 23:03:49
2
s_mn_05
s.m.n :
box steating, slight gain adjustment (turning down), and slight angle correction
2026-09-21 19:23:51
1
fiurelapolan
Fiury Matos :
i would lower the baseline or increase the scale
2026-09-21 20:43:28
1
uzuliz
Lizuzu :
Maybe increase wall filter to clean some noise
2026-09-21 03:28:56
3
nanakusi308
Eins_tein🧠 :
The sample volume needs to be reduced and adjusted to the center-most part of the vessel, this is to allow for correct sampling of the high velocities and laminar flow within the vessel. This will correct the retrograde appearance of the end diastolic wave.
2026-09-21 06:30:34
3
user3126078185085
user3126078185085 :
To measure pressure the cursor should be parallel to the flow
2026-09-21 07:48:43
2
michelleyuhasz
Michelle Yuhasz :
close sample volume for precose arteral measurements and l angle sample parallel to vessel walls
2026-09-21 21:50:56
1
mad_man2021
Manny m :
Box
2026-09-27 02:18:43
1
blacklake15
@blacklake :
As above but probably also reduce the PW gain.
2026-09-21 23:27:49
1
doritocrunchrap
doritocrunchrap :
heal-toe to get the vessel to be at a little more of an angle, fix the angle to my new angle and steer the box accordingly. probably turn down the doppler a little to clean up the noise.
2026-09-21 04:26:13
2
b.kdb
Badr :
the box angle should be steering with the vessel
2026-09-21 00:51:43
1
champ_dee_4.0
❤️Dee_Champ❤️ :
Steer the box and angle the cursor correctly.
2026-09-21 17:51:12
0
itsjustloli
Lorraine :
Sample volume
2026-09-23 22:32:05
0
bdw2017
M Hudson :
🥰🥰🥰
2026-09-22 04:19:54
0
To see more videos from user @ultrasoundtok, please go to the Tikwm homepage.